4. Au in vacuum
Dataset Spectra
No
Element; Emission line;
Photon energy 1486.61eV
Spectrum type Au 4f
Resolution 20
Dwell Time 1
Step size 0.1
Number of scans 1
Sample Metadata
Au
C
Ar
foil
Name 4. Au in vacuum
Ex situ preparation Clamped sample to stainless steel holder using stainless steel clips.
Additional Information Measured from a Au foil in vacuum, with the sample at a distance of 0.3 mm away from the nozzle. Nozzle diameter was 0.3 mm.
Dataset Metadata
Measurement Date 29-08-2020
Measurement conditions
Condition Value Units
pressure 1E-8 mbar
temperature 25 C
Spectrometer Settings
Analyzer Mode FAT
Area 0.3 mm diameter
Strength 100 W, 15 kV
Source Beam Size 0.3 mm diameter
Take Off Angle 90 degrees
Analyzer Acceptance Angle 22 degrees
Contributor Info
Laboratory Surface Analytics
Instrument Phoibos NAP-150
Contributors
Profile picture of Mark Greiner
Mark Greiner
Owner
Calibration Info
Protocol ISO 15472:2010
Calibration date 17-08-2020
Calibration value
Binding Energy 83.96
Emission line Au 4f7/2
Instrument value 1
Binding Energy 932.62
Emission line Cu 2p3/2
Suggested Tags

 

Au 00 0
No comments added

4. Au in vacuum 0
No comments added