Photon energy | 1486.61eV |
Spectrum type | survey |
Resolution | 20 |
Dwell Time | 0.9 |
Step size | 0.5 |
Number of scans | 1 |
Name | 1. Ag in vacuum |
Ex situ preparation | Sample clamped to stainless steel holder using stainless steel clamps |
In situ preparation | Sputter cleaned for 15 min using Ar at 3 kV and a pressure of 1E-5 mbar |
Additional Information | Spectra from a Ag foil measured in vacuum. Distance of sample to nozzle was 0.8 mm. Nozzle diameter was 0.8 mm. |
Measurement Date | 06-03-2020 |
Condition | Value | Units |
---|---|---|
pressure | 1E-8 | mbar |
temperature | 25 | C |
Analyzer Mode | FAT |
Area | 0.8 mm diameter |
Strength | 150 W, 15 kV |
Source Beam Size | 0.8 mm diameter |
Take Off Angle | 90 degrees |
Analyzer Acceptance Angle | 22 degrees |
Laboratory | Surface Analytics |
Instrument | Phoibos NAP-150 |
Protocol | ISO 15472:2010 |
Calibration date | 17-08-2020 |
Binding Energy | 83.96 |
Emission line | Au 4f7/2 |
Binding Energy | 932.62 |
Emission line | Cu 2p3/2 |