1. Ag in 2 mbar H2
Dataset Spectra
Photon energy 1486.61eV
Spectrum type survey
Resolution 20
Dwell Time 1.9
Step size 0.5
Number of scans 1
Sample Metadata
Ag
C
Ar
O
foil
Name 1. Ag in 2 mbar H2
Ex situ preparation Sample mounted on a stainless steel holder using stainless steel clips
In situ preparation Sputter cleaned for 15 minutes using 3 kV Ar at a pressure of 1E-5 and 10 mA emission current
Additional Information Ag foil measured in 2 mbar H2. Sample-to-nozzle distance was 0.8 mm. Nozzle diameter was 0.8 mm.
Dataset Metadata
Measurement Date 13-03-2020
Measurement conditions
Condition Value Units
pressure 2 mbar
ambient composition H2 none
temperature 25 C
Spectrometer Settings
Analyzer Mode FAT
Area 0.8 mm diammeter
Strength 150 W, 15 kV
Source Beam Size 0.8 mm diameter
Take Off Angle 90 degrees
Analyzer Acceptance Angle 22 degrees
Contributor Info
Laboratory Surface Analytics
Instrument Phoibos NAP-150
Contributors
Profile picture of Mark Greiner
Mark Greiner
Owner
Calibration Info
Protocol ISO 15472:2010
Calibration date 06-03-2020
Calibration value
Binding Energy 83.95
Emission line Au 4f7/2
Suggested Tags

 

Ag 00 0
No comments added
C 00 0
No comments added
Ar 00 0
No comments added

1. Ag in 2 mbar H2 0
No comments added