Photon energy | 1486.61eV |
Spectrum type | Sn 3d |
Resolution | 20 |
Dwell Time | 0.1 |
Step size | 0.05 |
Number of scans | 5 |
Name | 1. SnO2 in 0.1 mbar ethyl-glycol at 150 C, t=0-7min. |
Additional Information | Sn foil, with native oxide, at the beginning of heating in ethylene glycol. |
Measurement Date | 02-04-2196 |
Condition | Value | Units |
---|---|---|
pressure | 0.1 | mbar |
ambient composition | ethylene glycol | none |
temperature | 150 | C |
Analyzer Mode | FAT |
Area | 0.8 mm diameter |
Strength | 150 W |
Source Beam Size | 0.8 mm diameter |
Take Off Angle | 90 |
Analyzer Acceptance Angle | 22 |
Laboratory | Surface Analytics |
Instrument | Phoibos NAP-150 |
Protocol | ISO 15472:2010 |
Calibration date | 28-02-2020 |
Binding Energy | 83.96 |
Emission line | Au 4f7/2 |
Binding Energy | 932.64 |
Emission line | Cu 2p3/2 |
Binding Energy | 368.22 |
Emission line | Ag 3d5/2 |