Photon energy | 1486.61eV |
Spectrum type | survey |
Resolution | 100 |
Lens Mode | angle resolved |
Dwell Time | 0.1 |
Step size | 0.5 |
Number of scans | 1 |
Name | 1. Cu in vacuum |
Ex situ preparation | Sample mounted on a stainless steel holder using stainless steel clips. |
In situ preparation | Sputter cleaned for 15 minutes using 3 kV Ar at a pressure of 1E-5 and 10 mA emission current. |
Additional Information | Cu foil measured in vacuum after having been measured in 2 mbar O2. |
Measurement Date | 09-02-2020 |
Condition | Value | Units |
---|---|---|
pressure | 1E-8 | mbar |
temperature | 25 | C |
Analyzer Mode | FAT |
Area | 0.8 mm diameter |
Strength | 150 W, 15 kV |
Source Beam Size | 0.8 mm diameter |
Take Off Angle | 90 degrees |
Analyzer Acceptance Angle | 22 degrees |
Laboratory | Surface Analytics |
Instrument | Phoibos NAP-150 |
Protocol | ISO 15472:2010 |
Calibration date | 06-02-2020 |
Binding Energy | 83.96 |
Emission line | Au 4f7/2 |
Binding Energy | 932.64 |
Emission line | Cu 2p3/2 |