Pd metal reference
Dataset Spectra
Photon energy 1486.61eV
Spectrum type Pd 3d
Resolution 20
Lens Mode ARM22
Dwell Time 0.1
Step size 0.1
Number of scans 9
Sample Metadata
Pd
foil
Pd metal reference spectra
Name Pd metal reference
Ex situ preparation Cut off a piece of a 0.1 mm thick polycristallline Pd foil (Alfa Aesar, 99.9% purity)
In situ preparation the Pd foil was cleaned by several Ar+ sputtering and annealing cycles (in 0.5 mbar O2) at 700°C to soak out impurities. The last step was a heating step in vacuum (700°C) for recrystallization
Additional Information The reference spectra were measured at 300°C in vacuum
Dataset Metadata
Measurement Date 13-02-2019
Measurement conditions
Condition Value Units
pressure 1E-8 mbar
Temperature 300 °C
Spectrometer Settings
Analyzer Mode FAT
Area 0,8
Strength 150W
Source Beam Size 0,8
Take Off Angle 90
Analyzer Acceptance Angle 22
Contributor Info
Laboratory Surface Analytics
Instrument Phoibos NAP-150
Contributors
Profile picture of Mark Greiner
Mark Greiner
Contributor
Profile picture of Caroline Hartwig
Caroline Hartwig
Owner
Calibration Info
Protocol ISO 15472:2010
Calibration date 14-01-2019
Calibration value
Binding Energy 368.29
Emission line Ag 3d5/2